| Model Number |
H7493941848350 |
| Medical Device Problem Codes |
Material Integrity Problem (2978); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: synergy xd mr 3.50 x 48mm stent delivery system (sds) was returned for analysis and the following attributes were examined: stent profile: a visual examination of the stent found mid-stent damage, with stent struts bunched.The undamaged crimped stent od (outer diameter) was measured within maximum crimped stent profile measurement specification.Balloon profile: the balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.Tip profile: a visual and microscopic examination of the bumper tip showed no signs of damage.Hypotube profile: a visual and tactile examination of the hypotube found multiple hypotube kinks at several locations along the hypotube shaft.Shaft polymer extrusion profile: a visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
Reportable based on device investigation completed on 21june2022.It was reported that tip damaged occurred.A 3.50 x 48mm synergy xd device tip was observed damaged during preparation.The device was not used in the procedure.However, device analysis revealed stent damage.
|
| |
|