| Model Number |
H403023 |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
When preparing for the procedure, the ep-4 was off and self test failed which resulted in the cancellation of the procedure as the troubleshooting did not resolve the issue.There was no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation were inconclusive since the device was not returned for analysis.It was reported by the field in the communication hub, that the failure was caused by a faulty usb cable.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined that no non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported incident was consistent with the reported faulty usb cable.
|
| |
|
Event or Problem Description
|
|
The usb cable was replaced and the issue was resolved.
|
| |
|