| Model Number |
1804275-28 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a heavily calcified, moderately tortuous, 99% stenosed lesion in the obtuse marginal (om) artery.A 2.75x28mm xience skypoint drug eluting stent (des) was advanced with resistance from the anatomy and could not reach the lesion to be treated.The des was removed, also with resistance from the anatomy, and a non-abbott device was used to successfully complete the procedure.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no other incidents.The investigation determined the reported difficulties appear to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Search Alerts/Recalls
|