• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 24672
Medical Device Problem Codes Deflation Problem (1149); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that deflation failure occurred.The 90% stenosed target lesion was located in moderately tortuous and heavily calcified superficial femoral artery.A 4.0 x 150, 135cm mustang balloon catheter was advanced for dilatation.However, during the procedure there were problems noticed with the catheter and its progress while advancing.The device was then removed; however, the physician found it was hard to re-enter the lesion as probably it was not fully deflated.The procedure was completed with another of same device.There were no patient complications reported and the patient condition was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: mustang was received for analysis.A visual and tactile examination identified that the balloon was tightly folded and had not been subject to positive pressure.No issues were noted which may have potentially contributed to the complaint incident.A visual and tactile examination identified no kinks or damage to the shaft of the device.A visual examination identified no damage to the tip of the device.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that deflation failure occurred.The 90% stenosed target lesion was located in moderately tortuous and heavily calcified superficial femoral artery.A 4.0 x 150, 135cm mustang balloon catheter was advanced for dilatation.However, during the procedure there were problems noticed with the catheter and its progress while advancing.The device was then removed; however, the physician found it was hard to re-enter the lesion as probably it was not fully deflated.The procedure was completed with another of same device.There were no patient complications reported and the patient condition was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14803181
Report Number2134265-2022-06594
Device Sequence Number4653511
Product Code LIT
UDI-Device Identifier08714729793694
UDI-Public08714729793694
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K103751
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/18/2024
Device Model Number24672
Device Catalogue Number24672
Device Lot Number0026829957
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/02/2022
Supplement Date Received by Manufacturer07/15/2022
Initial Report FDA Received Date06/24/2022
Supplement Report FDA Received Date08/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexMale
-
-