• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINNACLE 100 ACET CUP 56MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINNACLE 100 ACET CUP 56MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Model Number 1217-01-056
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/26/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Dr.Wm had a patient show up in his clinic with a disassociated pinnacle liner.This patient had to undergo a revision surgery where both the original cup and liner were revised (revision was done on (b)(6) 2022).This patient is an active male, and had his original surgery 6 months prior (i do not have the actual date).The dissociation occurred 6 months post-op, where prior to this incident he was doing excellent and back to daily living.The original surgery was done using a 100 series cup and a 36x56 neutral liner.The surgeon does a very thorough job at examining his liners intra-op, and was not pleased with this complication.This is all the information i have.Did the patient experience a post-op device malfunction? yes, if yes, please describe.Please see event description.Patient needed to be admitted for revision surgery , did the patient experience an adverse event such as infection, non-union, allergic reaction, osteoporosis, overloading, pain, degenerative diseases, bleeding or oozing? no, did the patient require revision surgery or hardware removal? unknown, if no, was there any additional medical intervention required such as x-rays, additional procedures, prescriptions? yes - revision surgery , patient status/ outcome / consequences , patient consequence description/was there a clinical outcome experienced by the patient (infection, inflammation, etc.)? revision, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: yes, if yes, describe please reference event description , is the patient part of a clinical study unknown, (b)(6).Device property of none, device in possession of none, (b)(6).Device property of none, device in possession of none by checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst.True.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.All available x-rays were reviewed, and the allegation can be confirmed.Based on the attached photographs there is evidence to confirm the disassociation.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE 100 ACET CUP 56MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14803842
Report Number1818910-2022-11768
Device Sequence Number11981649
Product Code LPH
UDI-Device Identifier10603295008576
UDI-Public10603295008576
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1217-01-056
Device Catalogue Number121701056
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/26/2022
Supplement Date Received by Manufacturer07/22/2022
Initial Report FDA Received Date06/24/2022
Supplement Report FDA Received Date07/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALTRX NEUT 36IDX56OD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexMale
Patient Weight70 KG
-
-