| Model Number |
8655 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event was estimated based on aware date and no date was provided.
|
| |
|
Event or Problem Description
|
|
It was reported that device detachment occurred.During a procedure involving an opticross imaging catheter, the catheter broke into multiple sections.The procedure was completed with another of the same device.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
B3 date of event was estimated based on aware date and no date was provided.The device was returned for analysis.Visual inspection revealed a kink at the sheath assembly and the anchor housing detached.Microscopic inspection noted pitting and degradation of the transducer housing consistent with saline damage which occurs post-use of the device and is not related to the original device failure.
|
| |
|
Event or Problem Description
|
|
It was reported that device detachment occurred.During a procedure involving an opticross imaging catheter, the catheter broke into multiple sections.The procedure was completed with another of the same device.There were no patient complications reported.
|
| |
|