| Model Number |
ORT100 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The hydraulic fluid leak was traced to the o-ring gasket between the hydraulic fluid reservoir and the hydraulic pump.The gasket had degraded over time and required replacement.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2022, during preventive maintenance an imris customer service eningineer observed a small leak of hydraulic fluid in the wall mounted control panel of the t2x operating room table.There was no patient involvement and the table passed functional testing during the preventive maintenance service.
|
| |
|