MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS ESHEATH INTRODUCER SET; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS ESHEATH INTRODUCER SET; INTRODUCER, CATHETER Back to Search Results
Model Number 914ESJ
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Thromboembolism (2654)
Event Date 05/26/2022
Event Type  Injury  
Event Description
As reported through the (b)(6) tavi registry, a 23mm sapien 3 valve was transfemorally deployed in the aortic annulus.Right after tavr, the patient developed left femoral artery thromboembolism and occlusion was observed.Blood flow was restored by thrombus aspiration and balloon angioplasty.As per follow up, since there was no resistance during sheath insertion and removing, it is thought that tortuosity may have been contributed to the event.The outcome was determined as recovered.
 
Manufacturer Narrative
Per the instructions for use (ifu), thrombosis and embolization of air, calcification or thrombus are potential adverse events associated with transcatheter aortic valve replacement and bioprosthetic heart valves.According to the mayo clinic, some causes for thromboembolism not associated with invasive procedures include advanced age, atherosclerosis, cancer, previous mi, heart failure, dm, htn, a sedentary lifestyle, certain medications, and smoking.It is the natural tendency of the body to form a clot on foreign objects in the vascular space.These patients are anticoagulated for the procedure and interventional best practices mandate meticulous wiping and flushing of the devices to prevent and/or remove the clot.The thv training manuals and ifu instruct the operator to administer heparin and maintain the act at = 250 sec.In this case, there was no allegation or indication a device malfunction contributed to this adverse event.The exact cause of the reported event was unable to be determined but may have been due to procedural factors and or patient factors not provided.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS ESHEATH INTRODUCER SET
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key14807483
MDR Text Key295067420
Report Number2015691-2022-06389
Device Sequence Number1
Product Code DYB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K200258
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number914ESJ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/30/2022
Initial Date FDA Received06/24/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age84 YR
Patient SexFemale
-
-