| Model Number |
AVEA |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Vyaire medical file identification: (b)(4).Vyaire medical was able to confirm the issue, and a vyaire field service representative (fsr) evaluated the device on-site and replaced the o2 (oxygen) cell, resolving the issue.Blending accuracy was checked, and everything was within specifications.Ran a performance check; the unit passed all tests and operates in accordance with oem (original equipment manufacturer) specifications.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer reported to vyaire medical that the avea ventilator was having a high fraction of inspired oxygen (fio2) alarm.Furthermore, there was no patient associated with the reported event.
|
| |
|