MAUDE Adverse Event Report: BD KIESTRA LAB AUTOMATION BD KIESTRA¿ INOQULA+¿ TLA; MICROBIAL SPECIMEN INOCULATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD KIESTRA LAB AUTOMATION BD KIESTRA¿ INOQULA+¿ TLA; MICROBIAL SPECIMEN INOCULATION Back to Search Results
Model Number 447213
Medical Device Problem Code Structural Problem (2506)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that while using bd kiestra¿ inoqula+¿ tla the barcoda lid fell.There was no report of impact.The following information was provided by the initial reporter: "it was reported by the customer that hood gas springs need replacement.".
 
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.Medical device expiration date: na.
 
Event or Problem Description
It was reported that while using bd kiestra¿ inoqula+¿ tla the barcoda lid fell.There was no report of impact.The following information was provided by the initial reporter: " it was reported by the customer that hood gas springs need replacement.".
 
Additional Manufacturer Narrative
H6 investigation summary : the customer reported on the bd kiestra inoqula tla (material # 447213 - serial # (b)(6).That the hood gas springs needed replacement.The barcoda hood did not stay open.The field service engineer first swapped the hood gas springs from the barcoda 2 to the barcoda 1 until the new springs arrived.The new gas springs (material number 446064) were installed and tested without issues found.A capa and situation analyses have been opened for the investigation of corrective actions leading to avoid risks of errors and user injury.The issue that the hood gas springs needed replacement was confirmed by the field service engineer.Bd has initiated further root cause investigation related to barcoda lid fall through corrective and preventative actions to avoid risks of errors and user injury.This complaint has been included in the capa for traceability of the issue.Design history record (dhr) review is not required for this complaint.The complaint was evaluated via other elements of the investigation.The results of this evaluation have not identified any new hazards, new risks, or specific trends.Bd quality will continue to closely monitor for trends associated with this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD KIESTRA¿ INOQULA+¿ TLA
Common Device Name
MICROBIAL SPECIMEN INOCULATION
Manufacturer (Section D)
BD KIESTRA LAB AUTOMATION
6 marconilaan
drachten
Manufacturer (Section G)
BD KIESTRA LAB AUTOMATION
6 marconilaan
drachten
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14808398
Report Number3010141591-2022-00009
Device Sequence Number4878632
Product Code JTC
UDI-Device Identifier00382904472139
UDI-Public00382904472139
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number447213
Device Catalogue Number447213
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2022
Supplement Date Received by Manufacturer08/15/2022
Initial Report FDA Received Date06/24/2022
Supplement Report FDA Received Date08/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/28/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-