EXACTECH, INC. LOCKING FEMORAL IMPACTOR 0-5; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number LOCKING FEMORAL IMPACTOR 0-5
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
As reported, during impaction black plastic coating on the impaction handle chipped loose during impaction of the trial femur.Surgeon noticed right away and he checked the surgical site for debris.Confirmed clean and procedure carried on with no delay.There was no surgical delay/prolongation.Patient in stable condition.Sales rep was unable to obtain photos or x-rays.Product will be return.
 
Additional Manufacturer Narrative
Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
After further review of additional information received the following sections g3, g6, h2, h3 and h6 have been updated accordingly.(h3) based on historical complaint investigations and the returned device, the broken device reported was likely the result of years of surgical use and subsequent sterilization cycles, which resulted in a portion of the impaction face fracturing from the impactor body.The most probable root cause associated with the reported event of "fracture¿ is associated with a partial or full-thickness crack in the device materials.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING FEMORAL IMPACTOR 0-5
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key14823467
Report Number1038671-2022-00717
Device Sequence Number17296379
Product Code LXH
UDI-Device Identifier10885862059833
UDI-Public10885862059833
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLOCKING FEMORAL IMPACTOR 0-5
Device Catalogue Number213-64-01
Device Lot Number34534005
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2022
Supplement Date Received by Manufacturer07/25/2022
Initial Report FDA Received Date06/26/2022
Supplement Report FDA Received Date08/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-