ORTHALIGN, INC. LANTERN SURGICAL ASSISTANT; NAVIGATION UNIT Back to Search Results
Model Number 406000-01
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
During investigation of returned navigation unit the tester used an orthalign mechanically fabricated femoral platform, the navigational unit and in-house rs were attached and the femoral maneuver was performed three times.The error from the expected value was 0.5 degrees or less for each test in both flexion/extension and varus/valgus.An additional test on sawbones was performed to navigate the femoral resection plane to 3.5 degrees, as described in the complaint.A femoral maneuver was performed and the resection was dialed to 3.5 degrees of flexion.The orthalign engineer verified the image on the screen with the resection line matched the output.The engineer confirmed the cutting block was aimed in flexion, the software displayed flexion, and the resection line on the software image was also in flexion.A review of the log files revealed no anomalies or errors.The user did not register the angles they navigated to on the screen, so the target numbers were unable to be verified.A potential root cause for this issue could be improper femoral maneuver technique.This would explain the behavior seen by the user.The manuever must be performed as described in the surgical technique.A review of the device history record (dhr) was conducted.The device passed all manufacturing specifications prior to release.Orthalign, inc.Will continue to monitor this issue and take action if alert limits are exceeded.
 
Event or Problem Description
It was reported that orthalign flexion cut was going to be way off.Wanted four degrees.Was getting 18-20 degrees.So, we went to the manual im guide for distal femur and proximial tibia.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LANTERN SURGICAL ASSISTANT
Common Device Name
NAVIGATION UNIT
Manufacturer (Section D)
ORTHALIGN, INC.
120 columbia
suite 500
aliso viejo CA 92656
Manufacturer (Section G)
ORTHALIGN, INC.
120 columbia
suite 500
aliso viejo CA 92656
Manufacturer Contact
karyl haskell
120 columbia
suite 500
aliso viejo, CA 92656
MDR Report Key14824885
Report Number3007521480-2022-00006
Device Sequence Number14625156
Product Code OLO
UDI-Device Identifier00810832031406
UDI-Public00810832031406
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K200892
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number406000-01
Device Catalogue Number406000-01
Device Lot Number21111807LS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/29/2022
Initial Report FDA Received Date06/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-