MEDTRONIC IRELAND ENDURANT II EXTENSION STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
|
Back to Search Results |
|
| Model Number |
ETCF2323C49E |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A heli-fx endoanchoring system was used prophylactically with an endurant cuff in the endovascular treatment of a a pau in the distal aorta.A short distal landing zone distal of the pau to the aortic bifurcation was present and it was decided to use an aortic cuff due to the renal to bifurcation length. it was reported during the procedure, after placement of the aortic cuff the physician used the endoanchors to aid fixation on the graft at the proximal and distal landing zones.When theapplier was turned on the motor sounded very slow and different than normal.The endoanchor did load and it was checked to make sure the endoanchor was loaded.4 endoanchors were then placed at the top of the aortic cuff. the motor sounded very slow the entire time.Next the physician went to the distal aorta and placed the first endoanchor on the patients right side.When the physician went to place the second endoanchor, the physician went to stage one and the anchor went straight to deployment, it did not stop. the anchor was not aligned properly and went in at a steep angle.This caused the bottom of the stent graft to be pushed up.It was noted that the physician had not seen an applier to not stop at the first stage like normal.The physician said that he pushed the first stage button like normal and it just kept going.The physician then rotated the guide and went to the patients left side.It was then confirmed first that the endoanchor was loaded and it was, the physician then tested it to go to stage one and the anchor came out and then they reversed it to go back in.The physician at this point , placed the anchor in and when the physician went to go to stage one, the endoanchor went in and out with the motor making a clicking noise.The anchor at this point was stuck and the applier would not go forward or back. the endoanchor was stuck in the graft and applier.The physician gently rotated the device back and the anchor came dislodged from the graft, the physician then pulled the applier and guide out. a final angiogram was performed and the physician and saw no leak and the graft was noted to be slightly raised up on the patients right side.It was said that a cuff could be added if needed but it was then not required. per the physician the cause of the event was an issue with the applier. no additional clinical sequalae were provided and the patient is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis conclusion: the reported deployment difficulties and abnormal device response could not be assessed on the films provided; therefore, the cause of the events could not be fully determined.Lack of pre-implant ct¿s did not allow for a thorough assessment of the pre-implant anatomy.Angiogram videos showing the attempt to deploy the endoanchors was not shown on the films provided.A device issue cannot be ruled out.The malposition of the distal stent graft margin on the right side was confirmed and it is likely to have occurred during the deployment of the endoanchor.Analysis of the returned films did not reveal any endoanchor or stent graft integrity issues.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis conclusion: the reported deployment difficulties abnormal device response and malapposition events could not be assessed on the pre-implant films provided; therefore, the cause of the events could not be fully determined.Angiogram videos showing the attempt to deploy the e ndoanchors was not shown on the films provided.A device issue cannot be ruled out as a potential cause.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|