| Model Number |
DRILL SLEEVE, AVN |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2022, it was reported by a sales representative via email that an ar-3515h drill sleeve broke.This was discovered during a procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Search Alerts/Recalls
|