• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. LO-PRO SCRW,TI,3.0MMX 20MMCORT, STRL; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. LO-PRO SCRW,TI,3.0MMX 20MMCORT, STRL; SCREW, FIXATION, BONE Back to Search Results
Model Number LO-PRO SCRW,TI,3.0MMX 20MMCORT, STRL
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 06/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that during a mtp fusion surgery the cortical screw head sheared off when inserting.No excessive force was used.As the screw contacted the plate, one turn of the screwdriver caused it to shear off.They had to leave the broken screw in the bone.The surgery was finished successfully with the same device was used anyway.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
Complaint confirmed.Upon visual inspection, it was noted that only the head of the screw was returned for investigation.No problem was noted with the head of the screw.The cause of this remains undetermined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LO-PRO SCRW,TI,3.0MMX 20MMCORT, STRL
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key14828825
Report Number1220246-2022-05137
Device Sequence Number4813598
Product Code HWC
UDI-Device Identifier00888867210547
UDI-Public00888867210547
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K143614
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2024
Device Model NumberLO-PRO SCRW,TI,3.0MMX 20MMCORT, STRL
Device Catalogue NumberAR-8933-20S
Device Lot Number10505168
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2022
Supplement Date Received by Manufacturer06/08/2022
Initial Report FDA Received Date06/27/2022
Supplement Report FDA Received Date11/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/12/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-