• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY OPTIMA® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY OPTIMA® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE Back to Search Results
Model Number 787726
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that on (b)(6) 2022, the customer opened the package and removed the inlay optima ureteral stent after confirming that there was no abnormality with the outer package.During the process of delivering the pigtail straightener to the renal end, the stent was broken into three pieces.
 
Event or Problem Description
It was reported that on 07jun2022, the customer opened the package and removed the inlay optima ureteral stent after confirming that there was no abnormality with the outer package.During the process of delivering the pigtail straightener to the renal end, the stent was broken into three pieces.
 
Additional Manufacturer Narrative
The reported event is inconclusive because no sample was returned for evaluation.A potential root cause for this event could be, "material selection".As no patient involvement was reported the device was not used, however, it is intended for treatment purposes.It is unknown if the device had met all relevant specifications or contributed to the reported event.The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "avoid improper handling of stent such as bending, kinking, tearing, etc.Misuse could damage the overall integrity of the stent." "care should be exercised when removing the stent from the inner polybag to eliminate tearing or fragmentation." "the insertion of a ureteral stent should only be done by those individuals who have comprehensive training in the techniques and risks of the procedure." "directions for use: 1.Determine the proper stent length for the patient.This is generally calculated from the baseline pyelogram.Accurate measurements will optimize drainage efficiency and patient comfort.2.Insert the cystoscope then pass the guidewire* through the scope until the tip is in the renal pelvis.3.Move the pigtail straightener over the proximal end (kidney coil end) of the ureteral stent allowing easier insertion onto the guidewire.Remove pigtail straightener once the stent is secure on the guidewire.4.Pass the stent over the guidewire through the cystoscope by using the push catheter for proper placement." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY OPTIMA® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE
Common Device Name
URETERAL STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14829034
Report Number1018233-2022-04884
Device Sequence Number4652813
Product Code FAD
UDI-Device Identifier10801741015523
UDI-Public(01)10801741015523
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K043193
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/16/2024
Device Model Number787726
Device Catalogue Number787726
Device Lot NumberNGFR2781
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/07/2022
Supplement Date Received by Manufacturer11/30/2022
Initial Report FDA Received Date06/27/2022
Supplement Report FDA Received Date12/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age53 YR
Patient SexUnknown
Patient Weight80 KG
-
-