• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ADVANIX BILIARY; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ADVANIX BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00534690
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that an advanix biliary stent was used during a hemostasis procedure performed on (b)(6) 2022.During the procedure, it was noted that the handle could not break the stone.The procedure was successfully completed with another advanix biliary stent.There were no patient complications reported as a result of this event.The reported problem of handle could not break the stone does not match the intended use of the complaint device; therefore the exact complaint event is unknown.No further information has been obtained despite good faith efforts.Investigation results revealed that the stent was damaged, therefore this event has been deemed an mdr reportable event.
 
Additional Manufacturer Narrative
(b)(4).The returned advanix rx biliary preload stent was analyzed, and a visual evaluation noted that the push catheter and guide catheter did not present issues.The stent remained attached to the delivery system by the suture thread.The stent was observed damaged close to the proximal barb, and this could be what the customer reported.No other problem with the device were noted.The reported event was confirmed.Based on the provided and collected information, this device met all manufacturing requirements and no abnormalities were reported during the manufacturing process.After analysis, it was determined that the stent was damaged close to the proximal barb and this issue seems caused by tension, the push catheter and guide catheter did not present issues and during functional inspection the pull wire could be retracted and extended without issues.Therefore, based on the analysis performed and all the information gathered from the customer and the issues found the overall investigation conclusion code assigned will be "adverse event related to procedure" since the adverse event occurred during the procedure and the device had no influence on event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANIX BILIARY
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14830256
Report Number3005099803-2022-03396
Device Sequence Number19428659
Product Code FGE
UDI-Device Identifier08714729787242
UDI-Public08714729787242
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K101314
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/08/2023
Device Model NumberM00534690
Device Catalogue Number3469
Device Lot Number0028348585
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/26/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/30/2022
Initial Report FDA Received Date06/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-