MEDOS INTERNATIONAL SARL ENTERPRISE2 4MMX30MM; INTRACRANIAL NEUROVASCULAR STENT Back to Search Results
Model Number ENF403012
Medical Device Problem Code Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, and ethnicity were not provided.The name, and email address of the initial reporter are not available / reported.The initial reporter phone: (b)(6).[conclusion]: the healthcare professional reported that during an endovascular embolization procedure with the target at the internal carotid artery (ica), when the 4.0m dia x 30mm enterprise®2 stent (enf403012 / 6451023) was advancing through a prowler select plus 45 degree microcatheter, strong resistance was encountered at the microcatheter hub.The stent was fully retracted and removed without removing the microcatheter.The stent was replaced with another 30mm enterprise the device was reported to be without damage.On (b)(6) 2022, clarifying information was received.The information indicated that there was no report of any patient adverse event or patient complication as a result of the reported issue.The reported issue did not result in any clinically significant delay in the procedure.The procedure was successfully completed.The complaint device was returned for evaluation and analysis.The investigation is documented below.Investigation summary: a non-sterile 4.0m dia x 30mm enterprise®2 stent was received contained in a pouch.Visual inspection was performed.It was observed that the stent component was separated from the delivery wire inside the introducer.Also, the delivery wire was noted to be broken in two inside the introducer.The introducer was found in good normal conditions (i.E., no kinked or peeling conditions were observed on it).The stent was tried to be removed from the introducer, however, it could not be pushed or pulled due to the separated condition.The tip of the delivery wire was displacing the marker bands of the stent component.These conditions precluded proper functional testing, and no further inspection could be conducted on the stent component.The stent separation and delivery wire breakage as observed on the returned device could be secondary to the manipulation required to recapture the stent into the introducer.The stent and delivery wire components were returned inside the introducer, and the complaint documented that the stent was being advanced through the microcatheter.Although the exact timing of when the delivery wire breakage and separation from the stent could not be determined, it is most likely that these damages occurred during the post-operative handling/shipping of the device.The delivery wire separation from the stent would have caused the stent component to expand, which may have precluded the recapture into the introducer.Lake region medical performed a review of the device history records relative to the manufacturing, inspection, and packaging of the lot 6451023.The history record indicates this product was final inspection tested at lake region medical and was determined to be acceptable.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.The condition of the returned complaint device did not provide supportive evidence for the reported resistance issue encountered during the procedure, and functional test could not be performed due to the condition of the returned device, the reported issue documented in the complaint cannot be confirmed.There is no indication that the issue reported in the complaint is a result of a defect inherently related to the device.It should be noted that product failure could be caused by multiple factors.The instructions for use (ifu) does contain the following recommendations: if resistance is met during manipulation, determine the cause of resistance before proceeding.Do not apply undue force if resistance is encountered at any point during stent manipulation.Withdraw the unit and advance to a new one.If resistance is felt while recapturing the stent, do not continue to recapture the device.Withdraw the infusion catheter slightly to unsheathe the stent (without exceeding the recapture limit), and then attempt to recapture the stent again.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Event or Problem Description
The healthcare professional reported that during an endovascular embolization procedure with the target at the internal carotid artery (ica), when the 4.0m dia x 30mm enterprise®2 stent (enf403012 / 6451023) was advancing through a prowler select plus 45 degree microcatheter, strong resistance was encountered at the microcatheter hub.The stent was fully retracted and removed without removing the microcatheter.The stent was replaced with another 30mm enterprise.The device was reported to be without damage.Based on the result of the product analysis completed on (b)(6) 2022, the stent was returned separated from the delivery wire.The event has been deemed reportable as a "malfunction." on (b)(6) 2022, clarifying information was received.The information indicated that there was no report of any patient adverse event or patient complication as a result of the reported issue.The reported issue did not result in any clinically significant delay in the procedure.The procedure was successfully completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERPRISE2 4MMX30MM
Common Device Name
INTRACRANIAL NEUROVASCULAR STENT
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CODMAN AND SHURTLEFF, INC
325 paramount dr
raynham MA 02767
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
646591-7981
MDR Report Key14830341
Report Number3008114965-2022-00456
Device Sequence Number20032190
Product Code NJE
UDI-Device Identifier10886704075370
UDI-Public10886704075370
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
H60001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/28/2024
Device Model NumberENF403012
Device Catalogue NumberENF403012
Device Lot Number6451023
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2022
Initial Report FDA Received Date06/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-