It was reported that during a knee arthroscopy procedure using punch scissor straight, when the specialist was cutting the tendon, the forceps broke off.The broken pieces were removed from the patient with grasper forceps.The surgery did not depend on the instrument to be completed.No surgical delay nor further complications were reported.The patient's health condition is stable.
|
|
H10 h6 the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was no way to determine if the device contributed to the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.This case reports, while cutting with the punch scissor straight, the forceps broke off.Reportedly, the broken pieces were removed with grasper forceps.The device was not required to complete the procedure, and no procedural delay occurred.The patient¿s health condition is stable.Smith and nephew has not received the device and/or adequate materials to fully evaluate the complaint.If additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.The complaint was not confirmed.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, an impact event inconsistent with normal use, wear from prolonged use, misuse or rough handling, or inadequate routine maintenance.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|