| Model Number |
60-5274-032 |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Conmed japan reported on behalf of their customer that the 60-5274-032, spatula electrod w/stealth er 5mm x 32cm (5/cs) was being used during a laparoscopic cholecystectomy on an unknown date when it was reported that the ¿tip coating of electrode was peeling during the surgery.Seems that the hospital is presumed to be an initial defect, and it is unknown whether it came off during use with electricity (there is blood adhesion).¿ it was reported that there was no delay to surgery and that an unknown alternate device was used to complete the procedure.Further assessment found that the device did not fragment and did not fall into the surgical site.There was no report of injury, medical intervention, or hospitalization for the patient.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device is being returned to conmed for evaluation.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Additional Manufacturer Narrative
|
|
Received one 60-5274-032 in opened original package.Lot number was verified.Performed a visual inspection, the coating on the superior end of the electrode tip is peeling.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.A two-year lot history review was conducted and found no similar events reported for this lot number.A two-year review of complaint history revealed there has been a total of (b)(4) complaints, regarding (b)(4) devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised to examine the device prior to use for damage.Do not use if damage is found.These electrosurgical suction/irrigation instruments should only be used by surgeons trained in surgical endoscopy according to established hospital protocol.To avoid potential damage to trocar seals and to the electrosurgical tip, always slide the tip shield in the covered and locked position during insertion into the trocar cannula.Misuse of this or any other electrosurgical device can result in a patient injury.Read and become familiar with all instructions and directions before using this device.This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Event or Problem Description
|
|
Conmed japan reported on behalf of their customer that the 60-5274-032, spatula electrod w/stealth er 5mm x 32cm (5/cs) was being used during a laparoscopic cholecystectomy on an unknown date when it was reported that the ¿tip coating of electrode was peeling during the surgery.Seems that the hospital is presumed to be an initial defect, and it is unknown whether it came off during use with electricity (there is blood adhesion).¿ it was reported that there was no delay to surgery and that an unknown alternate device was used to complete the procedure.Further assessment found that the device did not fragment and did not fall into the surgical site.There was no report of injury, medical intervention, or hospitalization for the patient.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
|
| |
|