STRYKER TRAUMA KIEL LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø11X340MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 3425-1340S
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 04/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.A review of the device history is not possible because the lot number was not communicated.Should additional information become available, it will be provided in a supplemental report.Device disposition unknown.
 
Event or Problem Description
As reported: "breakage of the "gamma nail 11x340" with 85 shank screw inserted on (b)(6)2020.A revision surgery was performed.".
 
Additional Manufacturer Narrative
Please note correction to d9/h3 as the device was returned for evalaution.Please also note the corrections to h6 method, h6 results and h6 conclusion codes.The reported event could be confirmed, since the nail is broken apart as complained.The device inspection revealed the following: the visual inspection has shown the nail is broken apart at the lag screw hole.During the microscopic inspection the typical characteristics of a fatigue fracture could be identified on both fracture faces.The origin of the crack was at the lower corners of the bore, there is a fatigue zone with a smooth surface and progression lines over almost the whole surface of the fracture.At the the proximal fragment of the nail are very strong impression marks from the lag screw in the bore visible, which indicates that the nail was exposed to high loads over a long period of time.The relevant dimensions were verified during the investigation and no deviation from the specification could be detected.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to a patient factor related issue.The breakage of the device after 22 months of implantation, is most likely to be caused due to overloading of the device over a long period of time.Presence of signs of a fatigue fracture and absence of any remarkable user induced damage to the device supports this cause.In this relation following statement from the instructions for use can be mentioned: ¿¿ these devices can break when subjected to the increased loading associated with delayed unions and/or non-unions.Internal fixation devices are load sharing devices which are intended to hold fractured bone surfaces in apposition to facilitate healing.If healing is delayed or does not occur, the appliance may eventually break due to metal fatigue.Loads on the device produced by load bearing and the patient¿s activity level will dictate the longevity of the device." if more information is provided, the case will be reassessed.
 
Event or Problem Description
As reported: "breakage of the "gamma nail 11x340" with 85 shank screw inserted on (b)(6) 2020.A revision surgery was performed.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø11X340MM X 125°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14830698
Report Number0009610622-2022-00310
Device Sequence Number2367782
Product Code HSB
UDI-Device Identifier07613252273820
UDI-Public07613252273820
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K200869
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/31/2024
Device Model Number3425-1340S
Device Catalogue Number34251340S
Device Lot NumberK0D8584
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/02/2022
Supplement Date Received by Manufacturer09/14/2022
Initial Report FDA Received Date06/27/2022
Supplement Report FDA Received Date10/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/27/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
Patient SexFemale
-
-