Model Number 71655239
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code Pain (1994)
Date of Event 06/06/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a meta-nail tibial 11.5mm x 39cm had been implanted on an unspecified date, the patient started experiencing pain.A revision surgery was performed on (b)(6) 2022 while attempting to remove the nail, the nail cap was jammed to the nail and it could not be removed.The procedure lasted 6 hours until the nail was removed by ordinary methods.Patient has been discharged from the hospital and is currently doing fine.
 
Additional Manufacturer Narrative
Internal complaint reference: case-(b)(4).
 
Additional Manufacturer Narrative
The device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, without the requested clinical information, a thorough medical investigation cannot be rendered, nor could a definitive root cause of the reported failure be determined.Based on the information provided, the patient was revised due to pain.Per subsequent e-mail, it was reported, ¿the problem was in the tibial nail cap, it was like it sealed itself to the nail.¿ subsequently, it was reported the procedure lasted for 6 hours until the nail was removed by ordinary methods.According to the report, the patient is doing fine and discharged home.Since there was no harm has been alleged to this patient due to the prolonged use of anesthesia, no further clinical/medical assessment is warranted at this time.Should any additional relevant medical information be a review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of the instructions for use for this device reveal in the preoperative planning/surgical technique section, that correct surgical technique is essential to a successful outcome.This section also mentions to review the surgical technique for effective surgical procedures.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes could be alignment, fit/size of device used or wear.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
META-NAIL TIBIAL 11.5MM X 39CM
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14830993
Report Number1020279-2022-03122
Device Sequence Number2382422
Product Code JDS
UDI-Device Identifier03596010585707
UDI-Public03596010585707
Combination Product (Y/N)N
Initial Reporter CountrySA
PMA/510(K) Number
K061019
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71655239
Device Catalogue Number71655239
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/07/2022
Supplement Date Received by Manufacturer07/19/2022
Initial Report FDA Received Date06/27/2022
Supplement Report FDA Received Date07/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
META-NAIL NAIL CAP SET SCREW 0MM
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-