• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT LABORATORIES ARCHITECT CYCLOSPORINE WHOLE BLOOD PRECIPITATION REAGENT KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT LABORATORIES ARCHITECT CYCLOSPORINE WHOLE BLOOD PRECIPITATION REAGENT KIT Back to Search Results
Model Number 1L75-55
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer observed falsely decreased architect cyclosporine result which questioned by physician on one patient.The results provided were: initial=63 ng/ml /repeated=69 ng/ml /redrawn and repeated on same patient=170 ng/ml and 180 ng/ml /initial specimen repeated=69 ng/ml there was no reported impact to patient management.
 
Additional Manufacturer Narrative
The initial report was submitted under manufacturer report number 3008344661-2022-00066-01 abbott (b)(4) as manufacturing site)with suspect medical device of architect cyclosporine reagent kit, list 3r30.After further evaluation, the suspect medical device was changed to architect cyclosporine whole blood precipitation reagent kit, list 1l75 abbott (b)(4) as manufacturing site , which is under this report.
 
Additional Manufacturer Narrative
This follow-up submission being send to cross reference mdr 1415939-2022-00060-00 for likely cause changed on 21jul2022; to architect cyclosporine reagent kit, list number 01l75-25, lot number 37701fp00 from architect cyclosporine whole blood precipitation reagent kit, list 01l75-55, lot number 310275.After further evaluation it was determined that the cause of the issue was the architect cyclosporine reagent kit, list 01l75-25, lot number 37701fp00.) (abbottpark, il as manufacturing site) and submitted under manufacturer report number 1415939-2022-00060-00.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT CYCLOSPORINE WHOLE BLOOD PRECIPITATION REAGENT KIT
Common Device Name
CYCLOSPORINE
Manufacturer (Section D)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064
Manufacturer (Section G)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key14831877
Report Number1415939-2022-00047
Device Sequence Number4639892
Product Code MKW
UDI-Device Identifier00380740001322
UDI-Public00380740001322
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K080751
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1L75-55
Device Catalogue Number01L75-55
Device Lot Number310275
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/19/2022
Supplement Date Received by Manufacturer07/21/2022
Initial Report FDA Received Date06/27/2022
Supplement Report FDA Received Date08/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC I2000SR INST, 03M74-02, (B)(6) ; ARC I2000SR INST, 03M74-02, (B)(6)
Patient SexUnknown
-
-