• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PERIPHERAL ROTALINK PLUS; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PERIPHERAL ROTALINK PLUS; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 29720
Medical Device Problem Codes Contamination (1120); Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a device contamination occurred.A 1.50mm peripheral rotalink plus was selected for use.Upon receipt of the product, it was noted that there was a puncture to the box and it compromised the sterility of the device.There was no patient involvement.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the device was returned for evaluation.The advancer, handshake connections, sheath, coil, burr and annulus were visually and microscopically examined.Inspection of the device found that the outer and inner pouch appeared to be punctured.During review of the returned device, the manufacturing quality team was consulted and provided an explanation of the process requirements and inspection steps in place for the purpose of identifying damages of this nature.It was considered unlikely that the identified damages to the pouch occurred during manufacturing of the device.
 
Event or Problem Description
It was reported that a device contamination occurred.A 1.50mm peripheral rotalink plus was selected for use.Upon receipt of the product, it was noted that there was a puncture to the box and it compromised the sterility of the device.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERIPHERAL ROTALINK PLUS
Common Device Name
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14833564
Report Number2134265-2022-07297
Device Sequence Number11960688
Product Code MCW
UDI-Device Identifier08714729838906
UDI-Public08714729838906
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K121774
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/11/2024
Device Model Number29720
Device Catalogue Number29720
Device Lot Number0028856764
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/08/2022
Supplement Date Received by Manufacturer08/02/2022
Initial Report FDA Received Date06/27/2022
Supplement Report FDA Received Date08/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-