| Model Number |
29720 |
| Medical Device Problem Codes |
Contamination (1120); Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a device contamination occurred.A 1.50mm peripheral rotalink plus was selected for use.Upon receipt of the product, it was noted that there was a puncture to the box and it compromised the sterility of the device.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the device was returned for evaluation.The advancer, handshake connections, sheath, coil, burr and annulus were visually and microscopically examined.Inspection of the device found that the outer and inner pouch appeared to be punctured.During review of the returned device, the manufacturing quality team was consulted and provided an explanation of the process requirements and inspection steps in place for the purpose of identifying damages of this nature.It was considered unlikely that the identified damages to the pouch occurred during manufacturing of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that a device contamination occurred.A 1.50mm peripheral rotalink plus was selected for use.Upon receipt of the product, it was noted that there was a puncture to the box and it compromised the sterility of the device.There was no patient involvement.
|
| |
|