Model Number CNW0T6
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/31/2022
Type of Reportable Event Serious Injury
Event or Problem Description
A physician reported that following an intraocular lens (iol) implant procedure, a scratch was confirmed on the left side of the iol.The iol was removed and replaced with a backup lens during the initial procedure.Additional information has been requested.
 
Additional Manufacturer Narrative
A sample device was returned for analysis.Product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.Intraocular lens (iol) returned in four pieces (cut) and adhered to surgical gauze.Solution is dried on both surfaces of the optic and haptics.Both haptics are intact.The optic is torn/split-cut dividing the iol in four(4) and scratched/marked-rejectable.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Video shows intraocular lens (iol) implantation with company device accessories.The cartridge comes in picture and is partially filled with ophthalmic viscoelastic devices (insufficient amount of ovd).The lens is lifted from the lens case and positioned into the cartridge.The lens is advanced and implanted.A mark/line is visible over the iol optic.The mark appears to be on the posterior surface of the optic.The lens is cut and explanted.A scratch was observed over the posterior surface of the iol optic.The root cause may be related to a failure to follow the instructions for use (ifu).An inadequate amount of viscoelastic was observed in the cartridge.The ifu instructs to completely fill the cartridge with ovd (diagram provided) immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON TORIC IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14834346
Report Number9612169-2022-00295
Device Sequence Number14048857
Product Code MJP
UDI-Device Identifier00380652411226
UDI-Public00380652411226
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model NumberCNW0T6
Device Lot Number21288917
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/31/2022
Supplement Date Received by Manufacturer08/23/2022
Initial Report FDA Received Date06/27/2022
Supplement Report FDA Received Date09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III IOL, CARTRIDGE D; MONARCH III IOL, INJECTOR; PROVISC
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-