| Model Number |
MFS141 |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that 2 bd maxguard flow controller extension set with needleless y-site(s) were clogged.The following information was provided by the initial reporter: i have been told of 2 of the bd maxguard flow controller ext sets (dial-a-flows) not working at all.They are able to prime the controller extension but after priming no flow is allowed no matter the range that is selected.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary.Two samples were returned for investigation.The sets were examined for defects and abnormalities.No defects or abnormalities were observed.The sets were attached to a 10 ml bd syringe filled with saline and the flow controller in the open position.The set was unable to prime.The customer complaint that there is no flow was verified.A quality notification was sent to the manufacturer.The root cause is caused by a misplaced gasket.Corrective actions have been taken to correct the issue and the component that was received was constructed before the corrective actions were put in place.A device history record review for model mfs141 lot number 22039076 was performed.The search showed that a total of 11523 units in 1 lot number was built on 28jun2021.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.H3 other text : see h10.
|
| |
|
Event or Problem Description
|
|
It was reported that 2 bd maxguard flow controller extension set with needleless y-site(s) were clogged.The following information was provided by the initial reporter: i have been told of 2 of the bd maxguard flow controller ext sets (dial-a-flows) not working at all.They are able to prime the controller extension but after priming no flow is allowed no matter the range that is selected.
|
| |
|