• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® FOLEY CATHETER SILICONE COATED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® FOLEY CATHETER SILICONE COATED Back to Search Results
Model Number 123626A
Medical Device Problem Code Illegible Information (4050)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that new material code of bardia foley catheters did not have color coding on the packaging of the box (lot# mygn0175, lot# mycrrc36 and lot# myfv2209 ) and the wordings on the individual sterile packaging was too small (lot# mygn0175 , lot# mycrrc36 and lot# myfv2209 ).Per follow-up information received from ibc on 16jun2022, confirmed that one more lot number of bardia foley catheters did not had color coding on the packaging of the box (lot# myev2144 - ) and the wordings on the individual sterile packaging was too small (lot# myev2144 - ) based on the photo sample received on 09jun2022.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Additional Manufacturer Narrative
The reported event was unconfirmed because the device meets specifications.Received only photo of individual sterile packaging.Based on the attached photo, no abnormality was observed on the pouch print label as reported by the customer.The potential root cause for this failure mode could be due to illegible print & missing print.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "caution: this product contains natural rubber latex which may cause allergic reactions.Sterile: unless package has been opened or damaged.Warning: do not use ointments or lubricants having a petroleum base.They will damage latex and may burst balloon.Do not aspirate urine through the drainage funnel wall.Single use only.Do not desterilize.For urological use only.Valve type: use luer slip syringe.Do not use needle.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use syringe in the catheter valve.Never use more force than is required to make the syringe "stick" in the valve.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact an adequately trained professional for assistance, as directed by hospital protocol." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that new material code of bardia foley catheters did not have color coding on the packaging of the box (lot# mygn0175, lot# mycrrc36 and lot# myfv2209 ) and the wordings on the individual sterile packaging was too small (lot# mygn0175 , lot# mycrrc36 and lot# myfv2209 ).Per follow-up information received from ibc on 16jun2022, confirmed that one more lot number of bardia foley catheters did not had color coding on the packaging of the box (lot# myev2144 - ) and the wordings on the individual sterile packaging was too small (lot# myev2144 - ) based on the photo sample received on 09jun2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® FOLEY CATHETER SILICONE COATED
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14836278
Report Number1018233-2022-04927
Device Sequence Number4794263
Product Code KOD
UDI-Device Identifier00801741095245
UDI-Public(01)00801741095245
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K922431
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/01/2023
Device Model Number123626A
Device Catalogue Number0167V16S
Device Lot NumberMYCRRC36
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/09/2022
Supplement Date Received by Manufacturer09/14/2022
Initial Report FDA Received Date06/27/2022
Supplement Report FDA Received Date09/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/26/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-