• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL RUSCH EMERALD FO MACINTOSH BLADE; LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL RUSCH EMERALD FO MACINTOSH BLADE; LARYNGOSCOPE, RIGID Back to Search Results
Model Number IPN049940
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
Customer reported that: "the blades are getting stuck on the handle." no patient injury or consequence reported.Patient condition reported as "fine.".
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review could not be conducted since the lot number was not provided.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event or Problem Description
Customer reported that: "the blades are getting stuck on the handle".No patient injury or consequence reported.Patient condition reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).Further investigation was performed and the device was sent to the manufacturing site.The site has confirmed that the product is defective.The root cause was identified as manufacturing related.A non-conformance has been initiated to further investigate this issue.Teleflex will continue to monitor and trend on complaints of this nature.
 
Additional Manufacturer Narrative
(b)(4), the customer returned one emerald rusch mac 3 blade for evaluation.The blade was returned with a rusch greenspec handle attached.The blade was visually inspected for signs of misuse/abuse/damage.Nothing atypical was noted.A dimensional inspection could not be performed on the base of the blade due to the condition of the returned device.The mac 3 blade was functionally tested with the attached greenspec handle per ifu.The ifu states, "attach blade to handle.Click in place.To switch on, pull blade up.To switch off, move blade down." the blade was not able to be detached from the handle.The blade could not be removed even with significant force applied.R & d engineering was consulted as part of this investigation.R & d confirmed that the blade was defective as it could not be separated by manual force.The customer did not provide a lot number; therefore, a device history record review was performed based upon lot numbers from the sales history data of the customer.No relevant findings were identified.The complaint is confirmed.Functional testing was performed and the blade was not able to be detached from the handle.The blade could not be removed even with significant force applied.R & d engineering was consulted as part of this investigation.R & d confirmed that the blade was defective as it could not be separated by manual force.A dhr review was performed based on the sales history with no relevant findings identified.The sample has been sent to the manufacturing site for further evaluation.Teleflex will continue to monitor and trend on complaints of this nature.
 
Event or Problem Description
Customer reported that: "the blades are getting stuck on the handle".No patient injury or consequence reported.Patient condition reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUSCH EMERALD FO MACINTOSH BLADE
Common Device Name
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key14836519
Report Number3011137372-2022-00111
Device Sequence Number11350533
Product Code CCW
UDI-Device Identifier14026704640260
UDI-Public14026704640260
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberIPN049940
Device Catalogue Number004433300
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2022
Supplement Date Received by Manufacturer07/22/2022
07/27/2022
Initial Report FDA Received Date06/27/2022
Supplement Report FDA Received Date07/25/2022
07/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HANDLE.; HANDLE.; HANDLE.
Patient SexUnknown
-
-