| Model Number |
221270 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use, bd bbl¿ macconkey ii agar plates were contaminated.The following information was provided by the initial reporter: mat#221270 lot#2109300 plates have contamination growing inside.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6 investigation summary: this memo is to summarize findings regarding the complaint related to catalog number 221270, plate macconkey ii agar 100 ea, batch number 2109300 and bd complaint number 5404689 for contamination.During manufacturing of material 221270, media is formulated and sent through a high temperature short time sterilizer to remove bioburden.The petri dishes are subjected to uv radiation to decrease bioburden.The petri dishes are filled in a positive pressure hepa filtered environment.The filled plates are cooled and immediately wrapped into sleeves to decrease the introduction of microbes.Sleeves are then packaged into cartons and then transferred to a refrigerated truck (2 to 8 degrees c) for shipment to the distributor.Bd distributors are provided with the storage guidelines for the shipping and handling of bd media of 2 to 8 degrees c in a dark place.The batch history record for batch 2109300 was satisfactory and no quality notifications were generated during manufacturing and inspection.The release testing that is performed on this product does include bioburden testing.A sample of plates are incubated at 25 degrees c and at 35 degrees c for approximately 72 hours.All bioburden testing performed on this batch was satisfactory per bd internal procedures.Affected product does not have any sterility claims; the product is tested for bioburden prior to release to ensure that it conforms to product specifications.However, this does not ensure that the end-user will not receive a contaminated plate.Retention samples from batch 2109300 were not available for inspection.No return samples or photos were received for this investigation.The complaint history was reviewed, and one other complaint has been taken on batch 2109300 for contamination.Photos received for investigation of that complaint showed contamination in batch 2109300.This complaint can be confirmed for contamination.Bd has identified a contamination trend for this product and the investigation found opportunities for bioburden reduction in the manufacturing process.A capa (corrective and preventative actions) has been initiated and involves implementing additional cleaning events and evaluation of manufacturing procedures focused on in-process bioburden reduction.Additional trainings are planned with an ongoing training review for cleaning processes.Improvement in observation of contamination is expected as the capa progresses.Bd will continue to trend complaints for contamination.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use, bd bbl¿ macconkey ii agar plates were contaminated.The following information was provided by the initial reporter: mat#221270 lot#2109300 plates have contamination growing inside.
|
| |
|
Search Alerts/Recalls
|