SMITH & NEPHEW, INC. OXINIUM FEM HD 12/14 32MM -3; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERA/POLY, CEMEN OR NON-POROUS, UNCEM Back to Search Results
Model Number 71343203
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 02/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Case-(b)(4).
 
Event or Problem Description
It was reported that, after a thr was performed on (b)(6) 2019 the patient presented recurrent dislocation.Therefore, a revision surgery was performed on 23-feb-2022 in order to treat this adverse event, in which the oxinium fem hd 12/14 32mm -3 was removed and replaced with a new oxinium femoral head.The primary synergy stem was left in-situ.The surgeon does not blame the implants.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device did not confirm the stated failure mode.The device was found to be scratched and damaged along its surface, likely from the explantation process.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.The clinical/medical evaluation concluded that it was reported that a revision surgery was performed due to recurrent dislocation.The photos were reviewed, and visual inspection of the returned device noted, damage along base, likely from the event of the dislocation or the removal of the device from the patient; however, without supporting medical documentation, a thorough medical assessment of the clinical root cause of the dislocations cannot be performed.In the event additional medical/clinical records are received, the clinical task may be reopened, and a thorough assessment will be rendered at that time.It is noted, the surgeon does not blame the implants.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use document revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.The contribution of the device to the reported event could not be corroborated.Possible causes could include but not limited to traumatic injury, patient anatomy or abnormal loading of limb.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXINIUM FEM HD 12/14 32MM -3
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERA/POLY, CEMEN OR NON-POROUS, UNCEM
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14836693
Report Number1020279-2022-03135
Device Sequence Number17341804
Product Code LZO
UDI-Device Identifier03596010474216
UDI-Public03596010474216
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K021673
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71343203
Device Catalogue Number71343203
Device Lot Number19EM01472
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2022
Supplement Date Received by Manufacturer06/29/2022
Initial Report FDA Received Date06/27/2022
Supplement Report FDA Received Date06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/01/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age85 YR
Patient SexFemale
Patient Weight68 KG
-
-