| Model Number |
SIGLU60A |
| Medical Device Problem Codes |
Misfire (2532); Failure to Form Staple (2579)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Unspecified Tissue Injury (4559)
|
| Date of Event |
06/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic sleeve gastrectomy procedure, while stapling stomach, the knife did cut in the beginning without stapling.Then, cutting and stapling stopped, so there was a huge tissue damage, tissue loss, and bleeding.The staple line was incomplete at the distal.A new reload was opened, and the surgeon stapled near the damaged staple line to close it and stop bleeding.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: sigc60mt, tri 2.0 sigc60mt 60 med thk cartridge (lot#n1d0542y); sigphandle, sig power sigphandle handle (serial#unknown); sigpshell, sig power sigpshell control shell (lot#unknown); unk-sigadapt, unknown signia egia adapter (serial#unknown).Evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but two photos were available for evaluation.A video was also provided.Visual inspection of the video showed a surgical site presumably after a reload had been fired.Lots of bleeding could be seen in the image and it seemed as if the cut was incomplete.The two photos showed tissue outside of the patient.Malformed staples could also be seen in the photos.It was reported that staples were missing from the staple line after firing, the device initially fires, but failed to complete the firing cycle and the staples did not form as expected.The reported issues were confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.H3 evaluation summary: "medtronic conducted an investigation based upon all information received.The device and two photos were available for evaluation.A video was also provided.The visual inspection of the video showed a surgical site presumably after a reload had been fired.The visual inspection of the photos depicted lots of bleeding and it seemed as if the cut was incomplete.The two photos showed tissue outside of the patient.Malformed staples could be seen.Visual inspection of the returned device noted that the loading unit had damage to the cutting edge of the knife blade.During functional testing, the loading unit was loaded into a representative signia handle and adapter.The loading unit was loaded with a representative cartridge and was applied to test media.All staples were placed, and test media was transected.The interlock was tested and found to function properly.It was reported that there was poor staple formation.The reported issues was confirmed.The product analysis noted evidence that the device was not used as intended.It was also reported that there was partial firing and the staple line was incomplete.The reported issues could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly rev iewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: when positioning the stapler on the application site, ensure that no obstructions, such as clips, are incorporated into the instrument jaws.Firing over an obstruction may result in incomplete cutting action and/or improperly formed staples.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|