• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number DIU300
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Impairment (2138)
Date of Event 03/03/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device evaluation: the complaint lens was received inside of a specimen cup.Visual inspection under magnification revealed that the lens was received cut into three pieces.The lens was cleaned and lens damage separate from the cuts was observed.Based on the return condition of the lens, no further product evaluation could be performed.A product quality deficiency could not be determined.Based on the complaint investigation results, the complaint issue "unexpected postop refraction" was not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no other complaints were received for this production order.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.Attempts have been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the intraocular lens (iol) spherical equivalent (se) aim was emmetropia when target was residual myopia.Report received on mar(b)(6) ch 14, 2022, from a health care provider indicating they planed on explanting the j&j iol from the patient¿s left eye.The issue was noted one-week post-operative (op).Follow-up was done to confirm if the iol was explanted, but the customer did not respond.Upon completing the product investigation june 6 2022, it was noted that the iol had been returned and therefore explanted.Follow-up to confirm the explant date was done again, but the customer did not respond.No further information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14837805
Report Number3012236936-2022-01526
Device Sequence Number17303248
Product Code MJP
UDI-Device Identifier05050474747494
UDI-Public(01)05050474747494(17)240420
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 06/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/20/2024
Device Model NumberDIU300
Device Catalogue NumberDIU300U085
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/14/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2022
Initial Report FDA Received Date06/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-