Model Number 0010-0210R
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the assy refill bottle f fitting is empty.The procedure was not completed due to this event.There were no patient complications.This event is being reported for aborted/cancelled procedure with a patient under anesthesia.
 
Event or Problem Description
It was reported that the assy refill bottle f fitting is empty.The procedure was not completed due to this event.There were no patient complications.This event is being reported for aborted/cancelled procedure with a patient under anesthesia.
 
Additional Manufacturer Narrative
The console was not returned for analysis; however, this console was served at the customer's facility by a field service engineer (fse).The fse did not find any leaks with the di water refill bottle.No issues were found with the laser console.The console was calibrated and it passed full functional and electrical safety testing.Instructions were given to the customer on how to refill the di water into the refill bottle as it is outlined in the manual.Based on the information available, a conclusion code of no problem detected was assigned to this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GREENLIGHT XPS LASER SYSTEM
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
alyson harris
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key14838817
Report Number2124215-2022-23059
Device Sequence Number14660889
Product Code GEX
UDI-Device Identifier00878953001906
UDI-Public00878953001906
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K092735
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0010-0210R
Device Catalogue Number0010-0210R
Device Lot NumberXPS50292
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/16/2022
Supplement Date Received by Manufacturer08/11/2022
Initial Report FDA Received Date06/27/2022
Supplement Report FDA Received Date09/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/11/2011
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-