• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 36MM/+10; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 36MM/+10; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 6260-9-336
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/08/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported that the patient's right hip was revised.As reported by mps: "i covered a hip revision today due to trunnionosis failure in the right hip w/ dislocation (head dissociation)." spoke to mps.A stem, head and liner were revised.Explant picture for the liner shows damage on the rim.The mps cannot confirm if that damage was present in situ or if it is explant damage.
 
Additional Manufacturer Narrative
An event regarding disassociation involving a metal head was reported.The event was confirmed.Method & results -product evaluation and results: visual inspection: the device was not returned however photographs were provided for review.A review of the provided photographs indicated recently explanted devices covered in blood and tissue.Explantation damage observed on the insert.There is nothing remarkable on the head and stem.Material analysis, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: a review of medical records with a clinical consultant indicated "this inquiry presents a case of femoral head disassociation and reported trunnionosis.This process occurs between the trunnion of the femoral implant and the femoral head.As i mentioned above causes of this phenomenon are multifactorial including surgical technique, patient factors and implant factors.I can confirm that the event of disassociation occurred because i was able to review an x-ray which showed disassociation.I cannot confirm trunnionosis with the information provided.I cannot confirm the root cause for certain.The causes of disassociation are multifactorial as i noted above." -device history review: could not be performed as lot code information was not provided.-complaint history review: could not be performed as lot code information was not provided.Conclusion: it was reported the patient was revised due to disassociation of the head and stem.The device was not returned however photographs were provided for review.A review of the provided photographs indicated recently explanted devices covered in blood and tissue.Explantation damage observed on the insert.There is nothing remarkable on the head and stem.A review of medical records with a clinical consultant indicated "this inquiry presents a case of femoral head disassociation and reported trunnionosis.This process occurs between the trunnion of the femoral implant and the femoral head.As i mentioned above causes of this phenomenon are multifactorial including surgical technique, patient factors and implant factors.I can confirm that the event of disassociation occurred because i was able to review an x-ray which showed disassociation.I cannot confirm trunnionosis with the information provided.I cannot confirm the root cause for certain.The causes of disassociation are multifactorial as i noted above."no further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported that the patient's right hip was revised.As reported by mps: "i covered a hip revision today due to trunnionosis failure in the right hip w/ dislocation (head dissociation)." spoke to mps.A stem, head and liner were revised.Explant picture for the liner shows damage on the rim.The mps cannot confirm if that damage was present in situ or if it is explant damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V40 COCR LFIT HEAD 36MM/+10
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
brad curtis
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14839527
Report Number0002249697-2022-00934
Device Sequence Number4743856
Product Code JDI
UDI-Device Identifier07613327032321
UDI-Public07613327032321
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K022077
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number6260-9-336
Device Catalogue Number6260-9-336
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/08/2022
Supplement Date Received by Manufacturer09/02/2022
Initial Report FDA Received Date06/28/2022
Supplement Report FDA Received Date09/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age81 YR
Patient SexMale
-
-