Model Number 657310S
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.Should additional information become available, it will be provided in a supplemental report.
 
Event or Problem Description
As reported: "the locking screw did not lock to the plate.Spare screw was able to lock.".
 
Additional Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If more information is provided, the case will be reassessed.
 
Event or Problem Description
As reported: "the locking screw did not lock to the plate.Spare screw was able to lock.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW VARIAX 2 T10, FULL THREAD, 3.5MM / L10MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14840540
Report Number0008031020-2022-00291
Device Sequence Number4832661
Product Code HWC
UDI-Device Identifier07613327086799
UDI-Public07613327086799
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K132502
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Model Number657310S
Device Catalogue Number657310S
Device Lot NumberG31133
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/06/2022
Supplement Date Received by Manufacturer07/07/2022
Initial Report FDA Received Date06/28/2022
Supplement Report FDA Received Date08/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/25/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-