Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 05/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Device evaluated by mfr: customer has indicated that the product will not be returned to zimmer biomet for investigation, as the location of the product is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient had a two week follow up appointment where on x-ray it was found that an instrument was retained by the patient.Subsequently, the patient was revised and the instrument was explanted.
 
Additional Manufacturer Narrative
(b)(4).Reported event was unable to be confirmed due to limited information received from the customer.Device history record was reviewed and no discrepancies relevant to the reported event were found.The root cause of the reported issue is attributed to user not following instruction and left the instrument in the patient.Per the surgical technique which states that the patella protector should be part of the instrument count before closing the wound.It is not intended for implantation.Additionally, the instrument was etched with do not implant, indicating that the device should not be implanted/ left in the patient.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATELLA PROTECTOR 38 MM DIAMETER
Common Device Name
ORTHOPAEDIC PROSTHESIS IMPLANTATION INSTRUMENT / EXTREMITIES
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key14840686
Report Number0001822565-2022-01913
Device Sequence Number2350956
Product Code LXH
UDI-Device Identifier00889024217843
UDI-Public(01)00889024217843(10)64272837
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00598300138
Device Lot Number64272837
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2022
Supplement Date Received by Manufacturer09/14/2022
Initial Report FDA Received Date06/28/2022
Supplement Report FDA Received Date09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/06/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexFemale
-
-