| Model Number |
N/A |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
05/29/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant medical products: cpt 12/14 stem x-sml cocr; catalog#00-8114-040-00; lot#65179067, multipolar shell 55mm; catalog#5001-55; lot#64777093, multipolar liner 53/54/55 28mm; catalog#5001-53-28; lot#65177935, canal plug; catalog#8011-20-32; lot#65079750, report source: (b)(6).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was revised due to dislocation of the hip two days after implantation.No further patient outcome provided.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing that could be related to the reported event.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|