• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: I-SENS, INC. ASSURE PRISM MULTI BLOOD GLUCOSE TEST SYSTEM; SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

I-SENS, INC. ASSURE PRISM MULTI BLOOD GLUCOSE TEST SYSTEM; SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER Back to Search Results
Model Number 530001
Medical Device Problem Code High Readings (2459)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 09/23/2019
Type of Reportable Event Serious Injury
Event or Problem Description
On patient, customer received reading of 244 which seemed high to customer, so they had "rescue" (after further listening this is presumed to be the fire department) check the patient with their monitor, and it read 180."rescue" ended up taking the patient to the hospital.At the hospital the reading was 178.This all occurred with 20 minutes, so caller stated the patient couldn't have burned that much sugar.Caller is concerned that they will undertreat or overtreat if the readings on the meters aren't correct.Caller stated they just started using the assure prism meter system.Rep then checked variance of numbers.Variance is 26-27% and readings are all in zone a.Rep then offered to troubleshoot with the caller.Caller stated he tested controls with the meter, and they were in range.Caller stated the ranges are way too broad and don't reflect that a meter is as accurate as it should be.Rep then gathered caller's facility name, address, and phone number.The original call was regarding a patient who had overheated, was weak and dizzy and was a borderline diabetic.The medics checked and received the reading of 244.The patient hadn't eaten or taken any medication.The time between the readings of 244 and 180 was less than seven to eight minutes max according to the caller.Both tests were done via capillary testing, and caller is unsure if same finger was used for testing or not.Patient had been working on a roof in the heat and had become hot.The medics determined that patient was having hyperglycemia.Meters are kept in a jump bag in the ambulance in an air-conditioned climate-controlled environment.Patient's finger was cleansed with alcohol swab and allowed to dry before lancing finger.Test strips are stored in the original bottle and are not mixed.Caller then had questions about when to test with control solution and rep explained instances where control solution testing is encouraged.Rep then gathered meter serial number and strip lot information and explained replacement of meter, strips and return process.
 
Additional Manufacturer Narrative
Meter involved in incident was returned for investigation.Meter was evaluated and tested with retain strips of specified lot involved in incident.Products performed to specification.No failure detected.This mdr was submitted late due to technical issues with webtrader esg.Fda helpdesk provided assistance through fda esg helpdesk ticket number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURE PRISM MULTI BLOOD GLUCOSE TEST SYSTEM
Common Device Name
SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER
Manufacturer (Section D)
I-SENS, INC.
43, banpo-daero 28-gil
seocho-gu
seoul, 06646
KS  06646
MDR Report Key14840898
Report Number1832816-2022-00005
Device Sequence Number4640239
Product Code NBW
UDI-Device Identifier00015482530501
UDI-Public(01)00015482530013,(01)00015482530501
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/28/2022
Report Date (Section F) 06/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/20/2021
Device Model Number530001
Device Catalogue Number530001
Device Lot NumberRM21MAJ5D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date05/20/2022
Event Location Home
Date Report to Manufacturer06/28/2022
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date06/28/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient SexMale
Date Report Sent to FDA06/28/2022
-
-