|
On patient, customer received reading of 244 which seemed high to customer, so they had "rescue" (after further listening this is presumed to be the fire department) check the patient with their monitor, and it read 180."rescue" ended up taking the patient to the hospital.At the hospital the reading was 178.This all occurred with 20 minutes, so caller stated the patient couldn't have burned that much sugar.Caller is concerned that they will undertreat or overtreat if the readings on the meters aren't correct.Caller stated they just started using the assure prism meter system.Rep then checked variance of numbers.Variance is 26-27% and readings are all in zone a.Rep then offered to troubleshoot with the caller.Caller stated he tested controls with the meter, and they were in range.Caller stated the ranges are way too broad and don't reflect that a meter is as accurate as it should be.Rep then gathered caller's facility name, address, and phone number.The original call was regarding a patient who had overheated, was weak and dizzy and was a borderline diabetic.The medics checked and received the reading of 244.The patient hadn't eaten or taken any medication.The time between the readings of 244 and 180 was less than seven to eight minutes max according to the caller.Both tests were done via capillary testing, and caller is unsure if same finger was used for testing or not.Patient had been working on a roof in the heat and had become hot.The medics determined that patient was having hyperglycemia.Meters are kept in a jump bag in the ambulance in an air-conditioned climate-controlled environment.Patient's finger was cleansed with alcohol swab and allowed to dry before lancing finger.Test strips are stored in the original bottle and are not mixed.Caller then had questions about when to test with control solution and rep explained instances where control solution testing is encouraged.Rep then gathered meter serial number and strip lot information and explained replacement of meter, strips and return process.
|