• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. DRILL FLEXIBLE ENDOSCOPIC CANN 9MM; BIT, DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. DRILL FLEXIBLE ENDOSCOPIC CANN 9MM; BIT, DRILL Back to Search Results
Model Number 7209738
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/31/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during an acl reconstruction, femoral when drilling began in the femur, the flexible drill cannula broke.The drill bit was placed inside the 2.4mm guide, so it is drilled from the outside in with the 4.5mm bit, and this pushes the broken bit, then the doctor finishes removing it with a hemostat.To finish the procedure the 8.5mm flexible auger was used, from the same clancy system.Non-significant delay was reported.No further complications were reported.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was no relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Per case details, the broken drill was retrieved from the patient.The procedure was completed using a different device.No patient injuries or adverse consequences were reported.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.The complaint was not confirmed.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL FLEXIBLE ENDOSCOPIC CANN 9MM
Common Device Name
BIT, DRILL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14840935
Report Number1219602-2022-00918
Device Sequence Number4833220
Product Code HTW
UDI-Device Identifier03596010541994
UDI-Public03596010541994
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7209738
Device Catalogue Number7209738
Device Lot Number50421185
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/05/2022
Supplement Date Received by Manufacturer07/29/2022
Initial Report FDA Received Date06/28/2022
Supplement Report FDA Received Date08/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/21/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-