• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARGON MEDICAL DEVICES BIOPINCE ULTRA FULL CORE BIOPSY INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARGON MEDICAL DEVICES BIOPINCE ULTRA FULL CORE BIOPSY INSTRUMENT Back to Search Results
Model Number 360-1080-03
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The sample is indicated as available for return.As of the date of this report, the sample has not been returned.A follow-up report will be provided once the device has been received and reviewed.
 
Event or Problem Description
The throw adjuster is very heavy to move, and it is difficult to hit the exact point where it should stop.Moving the selector is not smooth, but choppy.The doctor thinks it is in place, but if it is slightly in front, it shoots 33 mm, even though the choice was meant to be 23 mm.Experienced very dangerous, and therefore the pbus that have a heavy button are discarded.
 
Event or Problem Description
1.The throw adjuster is very heavy to move, and it is difficult to hit the exact point where it should stop.Moving the selector is not smooth, but choppy.The doctor thinks it is in place, but if it is slightly in front, it shoots 33 mm, even though the choice was meant to be 23 mm.Experienced very dangerous, and therefore the pbus that have a heavy button are discarded.
 
Additional Manufacturer Narrative
The sample is indicated as returned.As of the date of this report, the sample has not been evaluated.A follow-up report will be provided once the device has been reviewed.
 
Event or Problem Description
1.The throw adjuster is very heavy to move, and it is difficult to hit the exact point where it should stop.Moving the selector is not smooth, but choppy the doctor thinks it is in place, but if it is slightly in front, it shoots 33 mm, even though the choice was meant to be 23 mm.Experienced very dangerous, and therefore the pbus that have a heavy button are discarded.
 
Additional Manufacturer Narrative
A review of the dhrs of the top level and sublots as well as the inspection records was conducted, and no deviations or non-conformances were found relating to this issue.There were two opened units returned by the customer for review.A visual in functional investigation was performed and it was found that the throw adjustor was extremely sticky and would not slide smoothly between length markers.This results in difficulty securing the throw adjustor to a set length and the potential for the device taking a longer specimen than desired when firing the biopince.The complaint is confirmed.An investigation into the sticky throw adjustor found that the most probable cause is an inadequate coating of teflon on the component, resulting in a significant increase of friction between the adjustor and the housing of the device.The operational procedure outlining the assembly of the biopince ultra devices was updated to develop a more robust teflon coating and final inspection process for the completed units.Training was provided to the operators to ensure proper adherence to the relevant procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOPINCE ULTRA FULL CORE BIOPSY INSTRUMENT
Common Device Name
BIOPINCE
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek rd
athens TX 75751
Manufacturer Contact
elnaz rahman
1445 flat creek rd
athens, TX 75751
9036759321
MDR Report Key14841359
Report Number0001625425-2022-01037
Device Sequence Number5245828
Product Code KNW
UDI-Device Identifier00886333224012
UDI-Public00886333224012
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K904987
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/09/2024
Device Model Number360-1080-03
Device Catalogue Number360-1080-03
Device Lot Number11372158
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2022
Supplement Date Received by Manufacturer06/17/2022
06/17/2022
Initial Report FDA Received Date06/28/2022
Supplement Report FDA Received Date07/11/2022
09/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-