This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified; however, the event most likely occurred due to the forceps channel becoming ruptured during user handling and material inside the endoscope escaped through the rupture.The event can be detected/prevented by following the instructions for use which state the following: "if a stream of air bubbles is emitted from the air/water nozzle even though the air/water valve is not being operated and the distal end of the insertion tube is 10 cm or more below the surface of the sterile water, there may be an irregularity in the air feeding function.If the endoscope is used while air is continuously fed, over-insufflation and patient injury may result.If air bubbles are emitted from the air/water nozzle, remove and reattach the air/water valve correctly, or replace it with a new one.If this fails to stop air bubbles from being emitted, do not use the endoscope, as there may be a malfunction.Contact olympus." " when inserting or withdrawing an endotherapy accessory, confirm that its distal end is closed or completely retracted into the sheath.Slowly insert or withdraw the endotherapy accessory straight into or from the slit of the biopsy valve.Otherwise, the biopsy valve or instrument channel may be damaged and pieces of it could fall off.It may cause patient injury." olympus will continue to monitor field performance for this device.
|