| Model Number |
TMICL13.2 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/16/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a 13.2mm tmicl13.2 implantable collamer lens, -11.50/+1.5/094 (sphere/cylinder/axis), in the patients right eye (od) on (b)(6) 2021.The lens was explanted on (b)(6) 2022 due to lens rotation not associated to a low vault.The lens was repositioned on (b)(6) 2021 but the problem was not resolved, so the lens was explanted.The lens was replaced with another same model/length lens and the problem was resolved.Post-op, the ucva was good but there was glare.The cause of the event was due to the device.The lens was discarded.
|
| |
|
Additional Manufacturer Narrative
|
|
Work order search: no similar complaint was reported for units within the same lot.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: b5- " the reporter indicated that the surgeon implanted a 13.2mm tmicl 13.2 implantable collamer lens of diopter-11.5/+1.5/094 (sphere/cylinder/axis) into the patient's right eye (od) on (b)(6) 2022.The patient experienced rotation not associated with a low vault.On (b)(6) 2021 the lens was rotated but this did not resolve the problem.On (b)(6) 2022 the lens was explanted and replaced with a lens of the same model/ length and the problem was resolved.The status of the eye was reported as "good ucva.C/o glare od (ticl) vs.Os (spherical)".The reporter indicated the cause of the event was the device and the device failed to perform as intended.The lens was discarded" claim# (b)(4).
|
| |
|
Search Alerts/Recalls
|