• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI ATTAIN STABILITY QUAD MRI SURESCAN; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) P

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI ATTAIN STABILITY QUAD MRI SURESCAN; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) P Back to Search Results
Model Number 479878
Device Problem Failure to Capture (1081)
Patient Problem Obstruction/Occlusion (2422)
Event Date 06/06/2022
Event Type  Injury  
Manufacturer Narrative
Continuation of concomitant medical products: 2088tc lead, implanted:(b)(6) 2018.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a competitor device change out procedure no capture was able to be obtained from the left ventricular (lv) lead in any configuration.The lead was capped and abandoned.It was noted that the lead was unable to be replaced at that time due to occlusion of the subclavian vein.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was further reported that a new lv lead was placed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN STABILITY QUAD MRI SURESCAN
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) P
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key14842657
MDR Text Key294873208
Report Number2649622-2022-12576
Device Sequence Number1
Product Code OJX
UDI-Device Identifier00643169860414
UDI-Public00643169860414
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P080006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/01/2021
Device Model Number479878
Device Catalogue Number479878
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/06/2022
Initial Date FDA Received06/28/2022
Supplement Dates Manufacturer Received11/04/2022
Supplement Dates FDA Received11/22/2022
Date Device Manufactured04/22/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
7120Q58 LEAD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age60 YR
Patient SexFemale
-
-