| Catalog Number |
05213312160 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Quality controls were performed on all of the involved meters and were acceptable.The customer stated the strips and meters are left in a car at times during health fairs and the strips are exposed to extreme temperatures at times.The meter and strips have been requested for return.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable cholesterol results from multiple accutrend plus meters.The customer used meter serial numbers (b)(4).The customer noticed they were getting more hi (>300 mg/dl) results with their accutrend meters than normal.The customer performed a comparison against cardiochek meters that provides a breakdown of the different cholesterol results and they got a hi on that brand.When they broke the result down, the total cholesterol result was 198 mg/dl.
|
| |
|
Additional Manufacturer Narrative
|
|
One vial of cholesterol test strips lot 566007-01 with 13 test strips inside and one vial of cholesterol lot 539009-21 also 13 test strips were received for investigation.No customer devices were available to be returned.Retention samples of lot 566007-01 and lot 539009-21 and the customer's returned strips were measured with low and high control solutions and 3 edta blood on reference meters.All results are within the range.No abnormalities were noticed with the returned strips.
|
| |
|