Model Number 20226
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/08/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that stent dislodgement occurred and the patient was sent to surgery.The target lesion was located in the mesenteric artery.A 8.0x20x135 cm express ld iliac/biliary stent was advanced for treatment.However, during the procedure, the stent was dislodged and the patient was sent to surgery to remove the device.There were no further patient complications reported.
 
Event or Problem Description
It was reported that stent dislodgement occurred and the patient was sent to surgery.The target lesion was located in the mesenteric artery.A 8.0x20x135 cm express ld iliac/biliary stent was advanced for treatment.However, during the procedure, the stent was dislodged and the patient was sent to surgery to remove the device.There were no further patient complications reported.It was further reported that the dislodged stent was pulled down to the common femoral artery and was retrieved intact by surgically opening the groin area.The patient condition post procedure was stable.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: a express ld device was returned for analysis.The balloon was received tightly folded with no stent attached.The stent was not returned with the device.A visual and tactile examination identified no kinks or damage to the shaft of the device.A visual examination of the device identified no issues with the markerbands or tip that could have contributed to the complaint incident.This concludes the product analysis.B2-outcomes attrib to adv event: removed other serious (important medical events) h6-patient codes: updated from insufficient information to no clinical signs, symptoms or conditions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS LD ILIAC / BILIARY
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14842813
Report Number2134265-2022-06963
Device Sequence Number2351011
Product Code NIO
UDI-Device Identifier08714729392460
UDI-Public08714729392460
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K133110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/30/2025
Device Model Number20226
Device Catalogue Number20226
Device Lot Number0028776732
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/08/2022
Supplement Date Received by Manufacturer07/20/2022
Initial Report FDA Received Date06/28/2022
Supplement Report FDA Received Date08/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexUnknown
-
-