The customer observed discrepant alinity c hemoglobin a1c results on multiple patient samples that were normal upon repeat on the alinity analyzer and variant (hplc).The following data was provided: (b)(6) 2022.Sid (b)(6) initial %hba1c result = > 14.0%, repeat = 5.70%, variant (hplc) = 6.20%, sid (b)(6) initial %hba1c result = > 14.0 %, repeat = 7.90%, variant (hplc) = 9.30%, sid (b)(6) initial %hba1c result = > 14.0%, repeat = 7.40%, variant (hplc) = 4.50%, sid (b)(6) initial %hba1c result = > 14.0%, repeat = 6.90%, variant (hplc) = 7.10%.(b)(6) 2022.Sid (b)(6) initial %hba1c result = > 14.0%, repeat = > 14%, variant (hplc) = 4.20%, sid (b)(6) initial %hba1c result = > 14.0%, repeat = 5.60%, variant (hplc) = 6.00%, sid (b)(6) initial %hba1c result = > 14.0%, repeat = 6.20%, variant (hplc) = 6.20%.(b)(6) 2022.Sid (b)(6) initial %hba1c result = > 14.0%, repeat = 6.00%, variant (hplc) = 6.40%, sid (b)(6) initial %hba1c result = > 14.0%, repeat = 8.00%, variant (hplc) = 7.80%, sid (b)(6) initial %hba1c result = > 14.0%, repeat = 7.30%, variant (hplc) = 7.40%, sid (b)(6) initial %hba1c result = > 14.0%, repeat = 6.60%, variant (hplc) = 6.60%, sid (b)(6) initial %hba1c result = > 14.0%, repeat = 6.30%, variant (hplc) = 6.30%, sid (b)(6) initial %hba1c result = > 14.0%, repeat = 10.30%, variant (hplc) = 10.20%, sid (b)(6) initial %hba1c result = > 14.0%, repeat = 6.00%, variant (hplc) = 6.40%, sid (b)(6) initial %hba1c result = > 14.0%, repeat = 8.00%, variant (hplc) = 7.80%, sid (b)(6) initial %hba1c result = > 14.0%, repeat = 8.50%, variant (hplc) = 8.50%.(b)(6) 2022.(b)(6) initial %hba1c result = > 14.0%, repeat = 5.30%, variant (hplc) = 5.50%, (b)(6) initial % hba1c result = > 14.0%, repeat = 5.70%, variant (hplc) = 6.00%, (b)(6) initial %hba1c result = > 14.0%, repeat = 4.80%, variant (hplc) = 4.80%, (b)(6) initial %hba1c result = > 14.0%, repeat = 7.70%, variant (hplc) = 7.50%, (b)(6) initial %hba1c result = > 14.0%, repeat = 5.00%, variant (hplc) = 5.10%.(b)(6) 2022.Sid (b)(6) %hba1c result = > 14.0%, sid (b)(6) %hba1c result = > 14.0%.No impact to patient management was reported.
|