C.R. BARD, INC. (COVINGTON) -1018233 SURESTEP INTERMITTENT CATHETER TRAY Back to Search Results
Model Number INTS14
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the silicone intermittent catheter were becoming disconnected during draining and nurse stated that the catheter was not being attached at all when they opened the package.It was also stated that the customer has observed the same with the pvc intermittent catheters.Also, nurse could not figure out how to empty the leg bag as they did not want to throw full bag of urine in the garbage, and it had occurred with two different nurses.The bag and the catheter were not together, so the nurse emptied the bladder with the catheter and collected it in the procedure tray.While the nurse had the catheter in the bladder draining urine, the leg bag came off the catheter in the middle of the procedure.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be "components out of specification".It is unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.A labeling review was not performed because labeling could not have prevented the reported failure.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that the silicone intermittent catheter were becoming disconnected during draining and nurse stated that the catheter was not being attached at all when they opened the package.It was also stated that the customer has observed the same with the pvc intermittent catheters.Also, nurse could not figure out how to empty the leg bag as they did not want to throw full bag of urine in the garbage, and it had occurred with two different nurses.The bag and the catheter were not together, so the nurse emptied the bladder with the catheter and collected it in the procedure tray.While the nurse had the catheter in the bladder draining urine, the leg bag came off the catheter in the middle of the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESTEP INTERMITTENT CATHETER TRAY
Common Device Name
INTERMITTENT CATHETER TRAY
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14843380
Report Number1018233-2022-04968
Device Sequence Number14041136
Product Code KOD
UDI-Device Identifier00801741183218
UDI-Public(01)00801741183218
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
K000723
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberINTS14
Device Catalogue NumberINTS14
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/08/2022
Supplement Date Received by Manufacturer10/25/2022
Initial Report FDA Received Date06/28/2022
Supplement Report FDA Received Date10/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-