MAUDE Adverse Event Report: ROCHE DIAGNOSTICS MG2; PHOTOMETRIC METHOD, MAGNESIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS MG2; PHOTOMETRIC METHOD, MAGNESIUM Back to Search Results
Catalog Number 06407358190
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation reviewed the customer's calibration results and there were calibration alarms noted on the date of the event.The investigation did find a successful calibration.The investigation reviewed the customer's qc results and the customer did have level 1 and level 2 results outside of the acceptable range.The investigation reviewed the system's alarm trace and found multiple abnormal aspiration alarms on the date of the event.This is an indication of poor sample quality.The field service engineer checked the instrument and did not find any abnormalities.The customer performed patient comparison testing with other analyzers and had acceptable results.Based on the provided information, the investigation determined the customer's actions of replacing the reagent pack resolved the issue.
 
Event or Problem Description
The initial reporter received questionable mg2 magnesium gen.2 results for nine patients tested on a cobas 8000 c 702 module serial number (b)(4).Prior to patient testing, the customer confirmed calibration and qc were within the acceptable ranges.The initial magnesium results were reported outside the laboratory.The customer performed qc and had results out of the acceptable ranges.The customer attempted to calibrate the reagent but had failing results.The customer removed the reagent pack and performed repeat testing with the samples on a different cobas 8000 c 702 module.Patient 1's initial magnesium result was 0.8 mg/dl.The patient's repeat magnesium result was 2.0 mg/dl.Patient 2's initial magnesium result was 0.7 mg/dl.The patient's repeat magnesium result was 1.8 mg/dl.Patient 3's initial magnesium result was 0.8 mg/dl.The patient's repeat magnesium result was 2.1 mg/dl.Patient 4's initial magnesium result was 1.2 mg/dl.The patient's repeat magnesium result was 3.2 mg/dl.Patient 5's initial magnesium result was 0.8 mg/dl.The patient's repeat magnesium result was 2.0 mg/dl.Patient 6's initial magnesium result was 0.8 mg/dl.The patient's repeat magnesium result was 2.1 mg/dl.Patient 7's initial magnesium result was 0.9 mg/dl.The patient's repeat magnesium result was 2.2 mg/dl.Patient 8's initial magnesium result was 0.8 mg/dl.The patient's repeat magnesium result was 2.1 mg/dl.Patient 9's initial magnesium result was 1.2 mg/dl.The patient's repeat magnesium result was 3.0 mg/dl.The customer determined the repeat results were correct.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MG2
Common Device Name
PHOTOMETRIC METHOD, MAGNESIUM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key14844059
Report Number1823260-2022-01895
Device Sequence Number17341860
Product Code JGJ
UDI-Device Identifier04015630930821
UDI-Public04015630930821
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K983416
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Catalogue Number06407358190
Device Lot Number62747901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/07/2022
Initial Report FDA Received Date06/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age77 YR
Patient SexFemale
-
-