• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24686
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The severely stenosed target lesion was located in an arteriovenous fistula.A 6.0 x80, 75cm gladiator elite balloon catheter was advanced for dilatation.However, during initial inflation at less working pressure for few seconds, the balloon ruptured.The device was removed and the procedure was completed with another of same device.No patient complications were reported and the patient status was stable.
 
Event or Problem Description
It was reported that balloon rupture occurred.The severely stenosed target lesion was located in an arteriovenous fistula.A 6.0 x80, 75cm gladiator elite balloon catheter was advanced for dilatation.However, during initial inflation at less working pressure for few seconds, the balloon ruptured.The device was removed and the procedure was completed with another of same device.No patient complications were reported and the patient status was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: gladiator elite 6.0 x80, 75cm was received for analysis.A visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.A balloon longitudinal tear was identified beginning approximately 4mm distal of the distal markerband and extending approximately 35mm proximally across the balloon material.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.The rated burst pressure for this device is 24 atmospheres.A visual examination of the markerbands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination found the shaft of the device to be kinked at more than one location.This type of damage is consistent with excessive force being applied to the device.A visual and tactile examination identified no issues or damage to the tip of the device.This concludes the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIATOR ELITE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14845245
Report Number2134265-2022-07447
Device Sequence Number4793940
Product Code LIT
UDI-Device Identifier08714729809753
UDI-Public08714729809753
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132810
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/03/2024
Device Model Number24686
Device Catalogue Number24686
Device Lot Number0028131754
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/21/2022
Supplement Date Received by Manufacturer08/04/2022
Initial Report FDA Received Date06/28/2022
Supplement Report FDA Received Date08/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age58 YR
Patient SexMale
Patient Weight75 KG
-
-