|
|
| Model Number |
GLJ-51-M |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic colectomy, at the time of anastomosis at the end of surgery, while doing a knotted suturing technique, the needle broke.The surgeon used a competitor's product to complete the procedure.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g1 (manufacturer name, mfr contact first name, last name, street 1, mfr city, region, postal code, email, phone number), g2, g3 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.One opened, one sealed samples and four photos were available for evaluation.Visual inspection noted the needle was noted to be broken.It was reported that the needle broke into separate pieces.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: the needle should always be held in the middle where the diameter is greatest.Grasping the needle near the swaged end or the tip could cause bends or breaks, or damage the connection between the needle and suture.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|